MECHATROVICH
QUALITY RESULTS INTERPRETATION

Gage R&R Acceptance Review

Sample case · One characteristic · Crossed study · Review question: Is this measurement system acceptable for preliminary process analysis?

Total Gage R&R6.93%
Below common 10% guideline*
Gage R&R / tolerance7.99%
Below common 10% guideline*

01 · Supported conclusion

The submitted aggregate Gage R&R result is below a commonly used 10% guideline. On that numerical basis, the system may be acceptable for the stated preliminary process-review purpose, subject to the organization’s own acceptance criteria, study design requirements, and intended decision risk.

02 · Interpretation risk

The operator means show a directional pattern with a 0.020 mm span. A passing aggregate percentage does not explain whether this pattern reflects technique, fixture interaction, part selection, resolution, or sampling noise.

Review flag: “Acceptable” should not be interpreted as “nothing further to investigate.” The operator pattern is small relative to the stated tolerance but still relevant to understanding the measurement method.

03 · Prioritized next checks

  1. Review the operator-by-part interaction and operator technique for systematic differences.
  2. Confirm that selected parts span the process variation the gage will encounter in normal use.
  3. Confirm resolution, fixture method, and work instruction consistency.
  4. Apply the organization’s approved acceptance criteria before release or customer submission.

04 · Boundary of this interpretation

This note interprets the sample output. It does not certify the gage, approve a customer submission, replace a complete MSA procedure, or provide final engineering approval on behalf of the client organization.